Summary
Type: SPE
Application #: 302000
Catalog#: 730012 - SPE Cartridges, Chromabond, C18 ec, 3ml, PP, 200mg, 50/pk
Detection: For further analysis we recommend HPLC with our column CC 125/3 NUCLEOSIL 100-5 C18 (Cat. No. 721622.30) Detailed information can be found in our
HPLC application 110580 'Analysis of malondialdehyde
Conditions: Column conditioning:
2 ml methanol, then 2 ml sodium phosphate buffer (0.125 mol, pH 3.0)
Sample application:
Slowly force or aspirate the solution from the sample pretreatment through the column
Column washing:
2 x 3 ml of a methanol - water (5:95, v/v); dry column under vacuum for 20 min.
Elution:
Coloured products are eluted with 800 μl methanol; extracts are concentrated in a stream of nitrogen
Authors: Suttnar J
Source: Analyt Biochemistry 249 (1997) 20-23